JOWA SEAGUARD

15 ppm 舱底水报警器(OWS 监测仪)

JOWA SEAGUARD is a 15 ppm oil content alarm on the discharge of the oily water separator (OWS), approved to IMO MEPC.107(49). It has two ppm alarm outputs, automatic sensor flushing with clean water and an event log on a sealed SD card (GMT date/time, ppm, separator status, valve, alarms, flow). The manual covers technical data, installation and setting the flow through the sensor, the menus (service, settings, log, test/simulation), downloading the log to a PC, maintenance, error codes, P&ID, drawings, spare parts and class certificates.

点击模型上的零件,查看编号和用途。打开 3D 说明书

保养规程

Weekly sensor cleaning and flushing

  1. Clean the sensor with the brush from the kit. 第 21 页
  2. Enter Service menu → Clean sensor. 第 21 页
  3. In the CLEAN SENSOR menu select Flush sensor (flush without zero calibration). 第 21 页

Monthly zero calibration

  1. Clean the sensor (see the weekly procedure). 第 21 页
  2. Make sure the water is completely clean and free of air bubbles. 第 21 页
    Reading before calibration about 0, maximum 2 ppm
  3. Service menu → Clean sensor → Zero calibration, confirm with YES. 第 21 页

Alarm check by simulation (MEPC.107(49) §4.2.5)

  1. Make sure the real value is below 15 ppm — otherwise the simulation will not start. 第 18 页
  2. Start the separator (otherwise the simulation is not recorded in the log). 第 19 页
  3. Test menu → Simulation, press YES. 第 19 页
  4. The reading rises from the real value to 20 ppm (2 s steps); at the set point the alarm trips and the overboard discharge valve closes. 第 19 页
  5. After returning to the real value the unit goes back to normal mode; a log entry is created. 第 19 页
    During the simulation the real value is still monitored — above 15 ppm a real alarm will trip

Downloading the log to a PC

  1. Connect a null-modem cable (2 × DB9 female) from the RS232 connector on the power board to the PC COM port; use a USB–RS232 adapter if needed. 第 16 页
  2. In a terminal program create a "Bilge alarm" connection, select the COM port, set 38400 baud, 8 bits, no parity, 1 stop bit, no flow control. 第 16 页
  3. Start capturing text to a file (Transfer → Capture text). 第 16 页
  4. On the unit: Log menu → List log files → enter the date → select the entry → Print to PC → format Bill or Sheet. 第 16 页
  5. Stop the capture in the terminal and open the saved file. 第 16 页

Setting the flow through the sensor

  1. Find the restrictor behind the coupling at the sample outlet (standard 1 mm orifice). 第 7 页
  2. If the flow is too low, drill the orifice 0.5 mm larger and check again; repeat in 0.5 mm steps. 第 7 页
    Nominal 2 l/min
  3. The restrictor must always be in the outlet before the coupling is connected. 第 7 页

保养周期

项目周期操作说明书
SensorWeeklyClean the sensor with the brush and run a manual flush (Service → Clean sensor → Flush sensor)第 21 页
SensorMonthlyZero calibration (Service → Clean sensor → Zero calibration) after cleaning, on clean water without bubbles; reading before calibration not more than 2 ppm第 21 页
Control unit batteryEvery 2 yearsReplace the battery inside the control unit第 21 页
Accuracy checkAt every renewal of the IOPP certificateCheck/calibration by the manufacturer or an authorised person, or replacement with a calibrated sensor; keep the certificate on board第 21 页

故障及处理方法

技术参数

Measuring range0–30 ppm (trend up to 40 ppm)第 5 页
Alarm 1 and 2 set points0–15 ppm, factory 15 ppm第 5 页
Alarm 1 / alarm 2 delay0–10 s / 0–540 s (factory 0)第 5 页
Power supply110–230 V AC 50/60 Hz, < 15 VA第 5 页
Sample and clean water pressure0.5–10 bar第 5 页
Sample flownominal 2 l/min第 5 页
Sample and ambient temperature+1…+60 °C第 5 页
Current output4–20 mA (or 0–20 mA) for 0–30 ppm, load up to 150 Ω第 5 页
Alarm contactsAlarm 1: 3 A/250 V; alarm 2: 3 A/24 V第 5 页
Pipe connectionsG ¼" female thread第 6 页
Response time to closing the overboard dischargenot more than 20 s (whole system)第 6 页
Restrictor at sample outletØ11 × 4 mm, 1 mm orifice第 7 页
Protection / weightIP65 / 7.5 kg第 5 页
Log to PC (RS232)38400 baud, 8 bits, no parity, 1 stop bit第 16 页

备件与目录号

零件编号数量说明书
Sensor unit86003-000011第 33 页
Flow restrictor86009-901131第 33 页
Power board86001-000301第 33 页
CPU board86001-000311第 33 页
SD card86001-000331第 33 页
3-way valve86001-000231第 33 页
Solenoid with contact86001-000241第 33 页
Cleaning brush83070-000021第 33 页
Fuse 1 A94007-52010第 33 页
Fuse 0.5 A94008-52050第 33 页
Flow switch19070-143861第 33 页
Battery86001-000361第 33 页
Printer88096-00020第 33 页
Null-modem cable88096-10001第 33 页
RS232–USB adapter88096-10002第 33 页
Sending the sensor to JOWA for calibration86010-00010第 21 页

来源:制造商说明书《Manual JOWA SEAGUARD》,共 51 页。页码以原文为准。作业前请核对船上的说明书。